Polymeric Liquid Membrane Electrodes Incorporated with Macrocyclic Hexaamines for Screening Adenine Nucleotides

Author(s): Iwona Szymanska, Hanna Radecka, Jerzy Radecki, Marek Pietraszkiewicz, Oksana Pietraszkiewicz

Journal Name: Combinatorial Chemistry & High Throughput Screening
Accelerated Technologies for Biotechnology, Bioassays, Medicinal Chemistry and Natural Products Research

Volume 3 , Issue 6 , 2000


Become EABM
Become Reviewer
Call for Editor

Abstract:

Lipophilic macrocyclic hexaamines supported by a poly(vinyl chloride) PVC matrix were used for the construction of liquid membrane electrodes sensitive toward adenine nucleotide polyanions. The membrane potential strongly depended on the pH of the sample solution. This phenomenon occurs due to the ability of the ionophore to accept protons. Therefore, the optimum pH was determined based on potential-pH profile. The potential measurements were carried out at pH 6.0 in the presence of 10-2 M 2- [N-morpholino] ethanesulfonic acid (MES) buffer. The potential response of these electrodes toward ATP- 4 and/or HATP-3 was close to the Nernstian slope. The selectivities against ADP-3, AMP-2, HPO4 -2 , and monovalent inorganic anions were estimated using the matched potential method. Chloride ions slightly affected potential response of the electrodes toward ATP-4/HATP-3. The influence of ionophore chemical structure on the selectivity and the sensitivity of these electrodes is briefly discussed.

Rights & PermissionsPrintExport Cite as

Article Details

VOLUME: 3
ISSUE: 6
Year: 2000
Page: [509 - 517]
Pages: 9
DOI: 10.2174/1386207003331436
Price: $65

Article Metrics

PDF: 2