Generic placeholder image

Current Pharmaceutical Analysis

Editor-in-Chief

ISSN (Print): 1573-4129
ISSN (Online): 1875-676X

Research Article

Development, Validation, and Application of a High-Performance Liquid Chromatography Method for the Analysis of Dissolution Samples of Telmisartan and Rosuvastatin Calcium Double-Layer Formulation

Author(s): Sang-Ho Lee, Sang-Ho Lee, Jong-Hyuk Lee* and Joo-Eun Kim*

Volume 17, Issue 10, 2021

Published on: 02 November, 2020

Page: [1282 - 1292] Pages: 11

DOI: 10.2174/1573412917999201102211941

Price: $65

Abstract

Background: Telmisartan and rosuvastatin calcium fixed-dose combination double-layer tablet is used for hypertension and hyperlipidemia treatment. Owing to the critical problem in establishing the specificity, precision, and accuracy of the USP analysis method and simultaneous analysis method for both components, an analysis method that could be applied to researchers in the field is urgently needed.

Objective: To develop and validate a reversed phase-high-performance liquid chromatography method for the quantitative analysis of dissolution samples of telmisartan and rosuvastatin calcium fixed-dose combination double-layer tablets, as a hypertension and hyperlipidemia treatment.

Methods: The developed analysis method was validated according to USP Category I requirements. The validity of the quantitative assay of the dissolution test solution was determined based on the system suitability, specificity, linearity, accuracy, precision, and solution stability of the assay for the components.

Results: The developed method was selective and precise. The retention times of telmisartan and rosuvastatin calcium was 8.1 min and 2.8 min, respectively, and system suitability was 0.217 and 0.17, respectively. Both components showed linearity at 10–120% concentration. The accuracy and precision were also within acceptable limits. The drugs in the dissolution samples were stable for a long time, no interaction occurred between the excipients and active pharmaceutical ingredients, and no interference was observed in the analysis.

Conclusion: The quantitative analysis of the double-layered formulation of telmisartan and rosuvastatin calcium was accurate, selective, and precise. Therefore, this method can be recommended for use in the industrial quality control of telmisartan and rosuvastatin calcium formulations.

Keywords: Telmisartan, rosuvastatin calcium, dissolution, method validation, quantitative analysis method, drug product, highperformance liquid chromatography.

Graphical Abstract

Rights & Permissions Print Cite
© 2024 Bentham Science Publishers | Privacy Policy