Title:Method to Determine the Delay Between Measurements in Two or More Spectrum Analyzers or Power Meters
VOLUME: 5 ISSUE: 2
Author(s):Luis C. Goncalves and Diogo F. da Cunha
Affiliation:Instituto de Engenharia Electrónica e Telemática de Aveiro, University of Aveiro, Aveiro, Marine Instruments S.A., Vigo
Keywords:Amplitude shift keying (ASK), direct current, fusion, power meter, spectrum analyzer (SA),
spectrum occupancy, trigger by hardware, trigger by software.
Abstract:The aim is to compute the time delay between the set of measurements in each of two or
more Spectrum Analyzers using a linear relation between two dimensions (power and time). This method
of time measurement implies to input in each Spectrum Analyzer two identical and synchronized
Amplitude Shift Keying signals, each one modulated by a square wave of duty cycle of 50%. It is also
valid with Direct Current. Instead of an input of an Amplitude Shift Keying signal (Radio Frequency
/Microwaves) a square wave is input and the power is measured with a Direct Current power meter with
trigger.