Abstract:

We proposed a method of synthesis to produce [11C]Choline using TRACERlab FXc module that utilized gas phase iodination. The product had radiochemical purity of 99.79 + 0.14 % and specific activity of 45.7 + 7.59 GBq/mol. [11C]Choline did not have at the moment a specific monograph in European Pharmacopeia therefore we used, when possible, as quality controls reference the monograph of [18F]FDG and we proposed suitable methods to verify radiochemical purity and to quantify residual DMAE and choline amounts.

Keywords: [11C]CH3I, [11C]Choline, gas phase iodination, prostate cancer, PET, quality controls, radiopharmaceutical, TRACERlab FXc